What is a Lateral Flow Immunoassay?
Lateral flow test strips based on the principles of immunochromatography are available for a wide range of target analytes. The first tests were performed for the detection of human chorionic gonadotropin (hCG). Today, there are commercially available tests for ovulation monitoring, infectious disease detection, drug abuse testing, and measurement of other analytes important to human physiology.
Products have also been introduced for testing, agricultural applications, environmental testing, and food and feed testing. While early tests presented qualitative results based on the presence or absence of a signal line, test design has progressed toward semi-quantitative and quantitative assays and the integration of handheld readers. A lateral flow immunoassay is described using variable terminology across different sectors and different countries.
Common names include:
- Lateral Flow Immunoassay (LFIA)
- Lateral Flow Test (LFT)
- Lateral Flow Device (LFD)
- Lateral Flow Assay (LFA)
- Lateral flow immunochromatographic assays
- rod
- quick test
- test strip
- quick test
Types Of Lateral Flow Tests
Lateral flow assays can be developed for use in a dipstick or cassette format. Both dipstick and cassette tests will perform similarly, it just depends on the industry, sample matrix, and market requirements, as to which format is suitable.
- Sandwich Format
The sandwich assay format is typically used to detect relatively large analytes. If the analyte has at least two distinct binding sites (i.e., epitopes), a “sandwich” assay can be developed in which an antibody against one epitope is conjugated to the nanoparticle and an antibody against another epitope is immobilized on the nanoparticle. test line. The sandwich format results in a signal intensity that is proportional to the amount of analyte present in the sample.
- Competitive Format
A competitive format is used to detect analytes where the analyte is too small for two antibodies to bind simultaneously, such as vitamins and antibiotics. In a competitive assay, the test line contains the target analyte molecule (usually a protein-analyte complex). The nanoparticles are conjugated to an antibody that recognizes the analyte. If the analyte is not present in the sample, the nanoparticle antibody conjugates will bind to the analyte in the test line, resulting in a high signal intensity.
If the target analyte is present in the sample, the analyte will bind to antibodies on the nanoparticle surface and prevent the nanoparticle from binding to the test line. This will reduce the signal on the test line resulting in a signal intensity that is inversely proportional to the amount of analyte present in the sample.
Traditional Side Flow Components
Lateral flow immunoassay technology uses a nitrocellulose membrane, coloured nanoparticles (or labels), and typically antibodies to produce results. When a sample is added, the sample will flow through the test device past the conjugate pad onto the nitrocellulose membrane and then onto the absorbent pad.
- Sample Pad – Serves as the first stage in the adsorption process, and in some cases contains a filter to ensure precise and controlled sample flow.
- Conjugate Pad: Stores the conjugate tags and antibodies, you will receive the sample. aids in the controlled release of the resolubilized conjugate on the nitrocellulose membrane.
- Nitrocellulose Membrane: Provides the ideal solid phase for immobilizing control and test line reagents. As the sample moves through the device, binding reagents located on the nitrocellulose membrane will bind to the target on the test line. A coloured line will form and the density of the line will vary depending on the amount of the target present.
- Absorbent Pad: Provides uniform capillary flow across the membrane, absorbs the applied sample and prevents backflow.
Sample Matrices
The target analyte and market requirements will determine the type of sample to be used in the assay. Some samples require an execution buffer to facilitate sample delivery, e.g. animal feed. Other samples, such as blood, serum, urine, or saliva, can be placed directly into a test, while there are times when a dilution buffer is required. Lateral flow immunoassays are developed to detect target analytes in sample matrices, including:
1. Milk
2. Whole blood
3. Serum
4. saliva
5. Urine
6. Fabric samples
7. food
8. Drink
9. Animal feed
10. Plant material
11. Water